Scale Calibration
Calibrating the Scale
Calibration at NO LOAD
Calibration at NO LOAD
Calibration at NO LOAD
Begin scale calibration upon successful entry into Calibration Mode.
Complete this calibration step within 30 seconds, or the calibration procedure exits.
Upon successful entry into Calibration Mode, the Scale Display blinks between CAL00 and CAL _ _, and the 7-segment display scrolls C00Lb, or C00g.
Ensure the platter is installed and there is no weighted load on it.
Touch the Scale Zero button on the front panel.
If Calibration at NO WEIGHTED LOAD is successful, continue to
Calibration at LOAD
.
If Calibration at NO WEIGHTED LOAD is not successful, see
Calibration Failure
.
Calibrating the Scale