If a Scale Display is unavailable, use the internal 7-segment one-character display to guide scale calibration.
If using a Scale Display during calibration, enable
Scale Display Configuration to enable the display and its port. Scale Display Configuration is disabled by default.
The MP72 scanner/scale must remain powered for at least 30 minutes after a cold power start before calibration.
At any time during calibration, if the units of measure are modified (for example, lb to kg) and a 30-second timeout completes, the Scale Display flashes CAL, and the 7-segment display has a U14 warning code (indicating Scale Out of Calibration).