Manual Entry into Calibration Mode

Manual Entry into Calibration Mode

For scales sold with a mechanical calibration switch installed, calibrate the scale manually. Some Weights and Measures regulatory jurisdictions require a mechanical calibration switch.
  1. Remove the platter (see Removing the Short or Medium Platter).
  2. If applicable, remove the security seal (paper labeled seal or tamper-evident film seal) over the calibration switch cover plug.
  3. Remove the calibration switch cover plug.
  4. Press and release the calibration switch (1).
    Scale/Calibration Switch
  5. Reinstall the platter and ensure there is no weighted load on the platter.
With successful entry into Calibration Mode, the scanner emits five long beeps. The Scale Display blinks CAL00 and CAL _ _ and the 7-segment display scrolls C00Lb or C00g. Continue to Program Legal Parameters.
If Calibration Mode was not entered successfully, the scale remains in its current state, or the Scale Display blinks CAL F. If the display blinks CAL F, correct the issue before re-entering calibration to enable an operational scale.