Audit Tallies

Audit Tallies

Audit Tallies used for scale verification (calibration counter = C, legal parameter counter = P, Legally Relevant firmware version number = F) are shown on the Scale Display or the 7-segment display.
  1. Press and hold the Scale Zero button for three seconds to access calibration tallies (not within Calibration mode). C###, P###, and #.##F tallies blink on the Scale Display and/or scroll on the 7-segment display.
    • C### represents the number of times the scale reached Calibration Success.
    • The P### represents the number of times any sealable parameter or legal parameter was changed with or without reaching Calibration Success.
    • #.##F represents the Legally Relevant firmware version number.
  2. Release the Scale Zero button to return the scale to normal operation and remove tallies from both displays.
    This feature is inhibited when the unit is in a scale Calibration Mode.
    Modifying the units of weight measure (for example, kg to lb) increments the parameter counter by two in the audit tally, because the Automatic Zero Setting legal parameter is enabled when changing from lb to kg, and disabled when changing from kg to lb.
  3. After verification, seal the scale by recording/reporting Audit Trail or apply physical seal to models that contain a physical calibration switch. Report information as required by local Weight & Measures Regulations, or where required by law.