Program Legal Parameters

Program Legal Parameters

The MP72 must be in Calibration Mode to optionally change the sealable and legal parameters (Unit Selection and Dampening Filter Setting). The Dampening Filter Setting can be changed multiple times during Calibration Mode without affecting the calibration outcome.
After completing any necessary changes, the calibration process re-starts with Calibration at NO LOAD. If you change the Unit Selection value before successfully reaching Calibration Success or Failure and the Calibration Mode exits, the scale remains non-operational until it is successfully calibrated.